Sign in →

Test Code FOL. FOLIC ACID

Important Note

Fasting specimen recommended.

Protect from LIGHT.

Additional Codes

FOLATE

N.FOLATE

Methodology

Automated Chemistry Analyzer (Alinity)

Specimen Requirements

Serum or Plasma

Container: SST tube or Green top Lithium heparin

Specimen Collection: 1 mL (minimum 0.5 mL)

  1. Allow specimen to clot completely if  a serum tube (SST) was used. Separate serum (SST) or plasma (Lithium Heparin) from cells within 24 hours and transport refrigerated. If red top tube is collected, separate serum from cells within 2 hours and place in separate plastic tube.
  2. Properly label tube with patients name, DOB, requisition number and type of tube, i.e. GREEN.
  3. PROTECT FROM LIGHT.

Specimen Transport/Stability

Refrigerated: 24 hours on gel, 7 days poured off protected from light

Specimen Rejection

1. Hemolyzed specimen

2. Non-light protected specimen

3. Improperly labeled specimen

Maximum Laboratory Time

24 to 48 hours

Reference Ranges

Folate      ng/mL
     7.0 - 31.4

 

Day(s) Test Set Up

Daily

Test Classification and CPT Coding

82746